Procedure – Powder/Chip Samples
Instrument calibration/drift correction.
Calibrate/drift following the procedure outlined in the operator's instruction manual.
Login Sample with the appropriate number of reps.
Weigh approximately *** grams of appropriately prepared sample into a *** Nickel Capsule, enter mass and sample identification into appropriate rep fields. If applicable, place the nickel capsule in the
appropriate autoloader position.
Repeat steps 3c through 3i for sample analysis.
氧氮氢分析仪的常见故障及解决方法
仪器启动时显示。没有水流。系统分析电流切断,分析停止。这是水流探测器不正常,水泵不工作,管道堵塞。将仪器的右面板取下,观察水流探测器,用手挤压补水塑料水瓶,发现水流正常,显示正常,证明水流探测器正常,管道畅通。启动循环水泵,但分析仪显示没有水流,此时可判定水泵不正常。打开水泵转子密封口,启动泵发现电机正常运转,此时关闭进水,拆下水泵,发现叶轮脱落。经了解,判定是由于外部冷却水停水,仪器内循环水温过高(水温应≤70℃),致使叶轮(叶轮材料PVC)热胀并脱轴。用粘合剂粘合叶轮后再粘于叶轮轴上,待粘合剂凝固后试车,仪器运行正常。 上海进口分析仪价格新型气路管道减少气路连接阀,使得漏气,断路故障更少。
Procedure – Powder/Chip Samples
Determine the instrument blank.
Press the Analyze button on the instrument screen, the lower electrode will close and the analysis sequence will start and end automatically.
Repeat steps 2b through 2g a minimum of three times.
Set the blank following the procedure outlined in the operator's instruction manual.
Instrument calibration/drift correction.
Login a minimum of three Standard reps.
Weigh approximately *** grams of a calibration/drift standard into a *** Nickel Capsule; enter the mass and standard identification into appropriate rep fields. If applicable, place the nickel capsule in the appropriate autoloader position.
Procedure - Solid Samples
Instrument calibration/drift correction.
Add approximately 0.05 g of *** Graphite Powder to a *** Graphite Crucible.
Firmly place the crucible on the lower electrode tip or appropriate autoloader position.
Press the Analyze button on the instrument screen, the lower electrode will close and the analysis sequence will start and end automatically.
Repeat steps 3b through 3j a minimum of three times for each calibration/drift standard used.
Calibrate/drift following the procedure outlined in the operator's instruction manual.
Method Selection
Two methods are described in this application note; either method can be used to analyze iron, steel, nickel-, and cobalt-base alloys. The Precision Method is recommended for general use and will provide the best precision and accuracy throughout the typical O, N, and H concentrations found in this group of metals; approximate cycle time is 3.5 minutes. The Fast Track method can be used where speed of analysis is a critical component; for example, when molten metal is being sampled and results are required in the shortest possible time. This method will produce suitable results for most samples; approximate cycle time is 2.25 minutes. As noted above, sampling and sample preparation are key elements to accurate O, N, and Hdetermination as well. It is up to the user to determine which method best meets their needs.
装有自动旁路阀的可快速拆装氧水捕集装置。上海进口分析仪价格
氧氮氢分析仪选哪家?给您推荐上海禹重!上海进口分析仪价格
力可OHN736&836系列仪器采用惰性熔融法,提供各种无机材料、钢铁、有色金属、硬质合金陶瓷材料中宽范围的氧氮氢同时测定。ONH836仪器通过***采纳客户的反馈意见和创新设计,力可***仪器采用触摸屏技术和在线软件平台,增强了仪器的可操作性,降低了分析成本,节省实验室桌面空间。
用户友好界面的Cornerstone™ 软件
力可**的Cornerstone碳硫分析软件采用触摸屏界面使用户可完整的达到分析控制、方法设定、在线诊断、数据报告。Cornerstone碳硫分析软件允许用户在单一分析界面中完成日常的所有工作,快速而且简单易用。力可创新的批次和重复样品数据分组功能极大简化了样品结果输出和结果标准偏差的自动计算。减少了繁琐的额外数据处理。