浙江氧氮氢分析仪厂家报价

时间:2021年10月27日 来源:

ONH836系列:氧氮氢分析仪(熔融法)

软件特点和优势

Cornerstone氧氮氢分析软件分成4个主要部分:分析、诊断、设定、仪器,界面可以快速导航和组织。工具条、滑动条和下拉式菜单简化了校正和数据处理的参数设定。分析软件同时包括实时的环境参数的监测,提供完整的系统图表模拟。高级诊断程序包括全数字化的在线操作手册、维护模拟演示、图片说明等提供快速指导。Cornerstone氧氮氢分析软件同时支持***的数据文档格式和可变的报告格式。 力可**的动态流量补偿控制DFC系统,保证了流量系统的完善,可以精确控制整个气路中的流量。浙江氧氮氢分析仪厂家报价

浙江氧氮氢分析仪厂家报价,分析仪

Procedure – Powder/Chip Samples

Instrument calibration/drift correction.

Calibrate/drift following the procedure outlined in the operator's instruction manual.


Analyze Samples.

Login Sample with the appropriate number of reps.

Weigh approximately *** grams of appropriately prepared sample into a *** Nickel Capsule, enter mass and sample identification into appropriate rep fields. If applicable, place the nickel capsule in the

appropriate autoloader position.

Repeat steps 3c through 3i for sample analysis.


浙江氧氮氢分析仪厂家报价工具条、滑动条和下拉式菜单简化了校正和数据处理的参数设定。

浙江氧氮氢分析仪厂家报价,分析仪

Procedure - Solid Samples

Determine the instrument blank.

Repeat steps 2b through 2g a minimum of three times.


Set the blank following the procedure outlined in the operator's instruction manual.


Instrument calibration/drift correction.

Login a minimum of 3 Standard replicates.


Weigh approximately *** grams of a calibration/drift standard, enter the mass and standard identification into appropriate replicate fields.



Note: LECO Reference Materia***o not require preparation.See preparation statement on the reference material certificate.





氧氮氢分析仪的常见故障及解决方法:

氧和氮空白值超过20。这是由于气流小,不能将炉子中的空气驱赶出去。可调节气体流量,调节载气压力在0.2~0.4MPa。接通仪器载气,放一个石墨坩埚在下电极上,打开主电源开关,点击软件上的关炉按钮,关闭炉子并等待10s。调节流量调节器,直到流量计a显示为30L/h,打开炉子。调节调节器直到流量计b显示为50L/h,再次关闭炉子。如果以上设置不稳定,则增加流量至100L/h,反复调节直至仪器稳定。 氧氮氢分析仪选哪家?给你推荐上海禹重科技!

浙江氧氮氢分析仪厂家报价,分析仪

Procedure – Powder/Chip Samples

Determine the instrument blank.


Press the Analyze button on the instrument screen, the lower electrode will close and the analysis sequence will start and end automatically.


Repeat steps 2b through 2g a minimum of three times.


Set the blank following the procedure outlined in the operator's instruction manual.


Instrument calibration/drift correction.

Login a minimum of three Standard reps.


Weigh approximately *** grams of a calibration/drift standard into a *** Nickel Capsule; enter the mass and standard identification into appropriate rep fields. If applicable, place the nickel capsule in the appropriate autoloader position.


Note: LECO Reference Materia***o not require preparation. See preparation statement on the reference material certificate. Solid Standards may be used to calibrate when chip or powder standards are not available.




***电极炉头设计,提高脉冲炉热交换效率。浙江氧氮氢分析仪厂家报价

通用下电极坩埚支座,容纳各种坩埚使用。浙江氧氮氢分析仪厂家报价

Method Selection


Two methods are described in this application note; either method can be used to analyze iron, steel, nickel-, and cobalt-base alloys. The Precision Method is recommended for general use and will provide the best precision and accuracy throughout the typical O, N, and H concentrations found in this group of metals; approximate cycle time is 3.5 minutes. The Fast Track method can be used where speed of analysis is a critical component; for example, when molten metal is being sampled and results are required in the shortest possible time. This method will produce suitable results for most samples; approximate cycle time is 2.25 minutes. As noted above, sampling and sample preparation are key elements to accurate O, N, and Hdetermination as well. It is up to the user to determine which method best meets their needs.








浙江氧氮氢分析仪厂家报价

信息来源于互联网 本站不为信息真实性负责